DEUTSCH
 ENGLISH
|
|

High-Power Microwave Metrology
Purpose Narrowband tubes (relativistic and non-relativistic), wideband and ultra-wideband system characterization
- HPM generator/source diagnostics:
measurement of current, voltage, local electric or magnetic field, etc.
- Free-field measurements
(electric and magnetic fields).
 HPM source diagnostics (Vircator driven by a Marx generator): measurement of several electrical parameters
Skills Probes and divider design, production and calibration at ISL
- Passive probes (high sensitivity and dynamics)
- H-field, E-field and current time derivative response: B-dot,
D-dot & I-dot probes
- direct time response: special current probe inside a
groove (up to 750 MHz) and TEM antenna (0.2 GHz–3 GHz)
- High-voltage carbon resistive divider with large bandwidth.
 Coaxial calibration cell
|
Facilities
- Calibration coaxial cell
- 170 mm diameter, cutoff frequency above 6 GHz
- CALMOPLA: 2 m x 2 m2 ground plane for sensor calibration
- good accuracy up to 6 GHz
- Electronic devices (fast oscilloscopes,
network analyzer, etc.)
- Anechoic chamber: 12.5 m x 5.5 m x 3 m
- absorption >27 dB [0.3–10 GHz]
- transmission <-90 dB [0.1–10 GHz].
 CALMOPLA: calibration with monopole on the ground plane
 B-dot calibration with CALMOPLA
 ISL anechoic chamber
Results
- Output characterization of fast high-power generator with 50 W carbon divider
(DC -1.7 GHz, up to 500 kV)
- Electron beam / plasma current measurement (several tens of kA)
- Free-field measurement beyond 100 kV/m from 100 MHz to 6 GHz.
|
|
|
|
|
|
|
|
 |
|
|
| |
 |
 |
|
|